Specifications – 2250 Light Platform
Specifications apply to 2250 Light fitted with Microphone Type 4950
and Microphone Preamplifier ZC-0032
SUPPLIED MICROPHONE
Type 4950: Prepolarized Free-field ½
Microphone
Nominal Open-circuit Sensitivity: 50 mV/Pa (corresponding to
26 dB re 1 V/Pa) ± 2 dB
Capacitance: 12.5 pF (at 250 Hz)
MICROPHONE PREAMPLIFIER ZC-0032
Nominal Preamplifier Attenuation: 0.3 dB
Connector: 10-pin LEMO
Extension Cables: Up to 100 m in length between the microphone
preamplifier and 2250 Light, without degradation of the specifications
SELF-GENERATED NOISE LEVEL
Typical values at 23°C for nominal microphone open-circuit
sensitivity:
KEYBOARD
Pushbuttons: 11 keys with backlight, optimised for measurement
control and screen navigation
ON-OFF BUTTON
Function: Press 1 s to turn on; press 1 s to enter standby; press for
more than 5 s to switch off
STATUS INDICATORS
LEDs: Red, amber and green
DISPLAY
Type: Transflective back-lit touch screen 240 320 dot matrix
Black and White Scheme
Backlight: Adjustable level and on-time
USER INTERFACE
Measurement Control: Using pushbuttons on keyboard
Setup and Display of Results: Using stylus on touch screen or
pushbuttons on keyboard
Lock: Keyboard and touch screen can be locked and unlocked
USB INTERFACE
USB 1.1 OTG Mini B socket
MODEM INTERFACE
Hayes compatible GSM or standard analogue modems connected
through the Compact Flash slot
HEADPHONE SOCKET
Connector: 3.5 mm Minijack stereo socket
Max. Peak Output Level: ±1.4V
Output Impedance: 32 in each channel
EXTERNAL DC POWER SUPPLY REQUIREMENTS
Used to charge the battery pack in the instrument
Voltage: 8 24 V DC, ripple voltage <20 mV
Current Requirement: min. 1.5 A
Power Consumption: < 2.5W, without battery charging, < 10 W
when charging
Cable Connector: LEMO Type FFA.00, positive at centre pin
BATTERY PACK
Type: Li-Ion rechargeable
Typical Operating Time: > 8 hours
STORAGE SYSTEM
Internal Flash-RAM (non-volatile): 20 Mbyte for user setups and
measurement data
External Secure Digital Memory Card (SD-card): For store/recall
of measurement data
External Compact Flash Memory Card (CF-card): For store/recall
of measurement data
CLOCK
Back-up battery powered clock. Drift < 0.45 s per 24 hour period
WARM-UP TIME
From Power Off: < 2 minutes
From Standby: < 10 seconds
TEMPERATURE
IEC 60068–2–1 & IEC 60068–2–2: Environmental Testing. Cold and
Dry Heat
Operating Temperature: –10 to + 50°C (14 to 122°F), < 0.1 dB
Storage Temperature: –25 to +70°C (–13 to +158°F)
HUMIDITY
IEC 60068–2–78: Damp Heat: 90% RH (non-condensing at 40°C
(104°F)).
Effect of Humidity: < 0.1 dB for 0% < RH < 90% (at 40°C (104°F)
and 1 kHz)
MECHANICAL
Environmental Protection: IP 44
Non-operating:
IEC 60068–2–6: Vibration: 0.3 mm, 20 m/s
2
, 10 500 Hz
IEC 60068–2–27: Shock: 1000 m/s
2
IEC 60068–2–29: Bump: 4000 bumps at 400 m/s
2
WEIGHT AND DIMENSIONS
650 g (23 oz.) including rechargeable battery
300 93 50 mm (11.8 3.7 1.9
) including preamplifier and
microphone
LANGUAGE
User Interface in Catalan, Croatian, Czech, Danish, English, Flemish,
French, German, Hungarian, Japanese, Italian, Polish, Portuguese,
Romanian, Serbian, Slovenian, Spanish, Swedish and Turkish
HELP
Concise context-sensitive help in Catalan, English, French, German,
Italian, Japanese, Polish, Portuguese, Romanian, Serbian, Slovenian
and Spanish
Weighting Microphone Electrical Total
“A” 14.0 dB 12.9 dB 16.4dB
“B” 12.9 dB 11.8 dB 15.4dB
“C” 13.0 dB 13.4 dB 16.2dB
“Z” 5 Hz–20 kHz 14.4 dB 19.2 dB 20.4dB
Compliance with Standards
,
CE-mark indicates compliance with the EMC Directive and Low Voltage Directive.
C-Tick mark indicates compliance with the EMC requirements of Australia and New Zealand.
Safety EN/IEC 610101: Safety requirements for electrical equipment for measurement, control and laboratory use.
UL 61010B1: Standard for Safety – Electrical measuring and test equipment.
EMC Emission EN/IEC 61000–6–3: Generic emission standard for residential, commercial and light industrial environments.
CISPR 22: Radio disturbance characteristics of information technology equipment. Class B Limits.
FCC Rules, Part 15: Complies with the limits for a Class B digital device.
IEC 61672–1, IEC 61260, IEC 60651 and IEC 60804: Instrumentation standards
EMC Immunity EN/IEC 61000–6–2: Generic standard – Immunity for industrial environments.
EN/IEC 61326: Electrical equipment for measurement, control and laboratory use – EMC requirements.
IEC 61672–1, IEC 61260, IEC 60651 and IEC 60804: Instrumentation standards
Zobrazit stránku 8