11
For Display and Storage:
Weather Data (requires connection to a weather station):
Wind Dir. avg.
Wind Dir. min.
Wind Dir. max.
Wind Speed avg.
Wind Speed min.
Wind Speed max.
Amb. Temperature
Amb. Humidity
Amb. Pressure
Amb. Rain Gauge
Only for Display as Numbers or Quasi-analog Bars
Instantaneous Weather Data:
Wind Dir.
Wind Speed
Instantaneous GPS Data:
Latitude
Longitude
MEASURING RANGES
When using Microphone Type 4950
Dynamic Range: From typical noise floor to max. level for a 1 kHz pure
tone signal, A-weighted: 16.4 to 140 dB
Primary Indicator Range: In accordance with IEC 60651, A-weighted:
23.6 dB to 122.3 dB
Linearity Range: In accordance with IEC 60804,
A-weighted: 21.5dB to 140.8dB
Linear Operating Range: In accordance with IEC 61672, A-weighted
1 kHz: 24.9 dB to 139.8 dB
Peak C Range: In accordance with IEC 61672: 1 kHz, 43.0 dB to
142.8 dB
SAMPLING FOR BROADBAND STATISTICS
Statistics can be based on either L
AF
, L
AS
or L
Aeq
:
•Statistics L
AFN1–7
or L
ASN1-7
are based on sampling L
AF
or L
AS
, resp.,
every 10 ms into 0.2 dB wide classes over 130 dB
•Statistics L
AN1–7
are based on sampling L
Aeq
every second into 0.2 dB
wide classes over 130 dB
Full distribution saved with measurement
The Std.Dev. (Standard Deviation) parameter is calculated from the
Statistics
Measurement Display and Control
MEASUREMENT DISPLAYS
Measurement data displayed as numbers of various sizes and one quasi-
analog bar.
Measured data are displayed as dB values, housekeeping data as
numbers in relevant format.
Instantaneous measurement L
XF
is displayed as a quasi-analog bar
MEASUREMENT CONTROL
Manual: Manually controlled single measurement
Automatic: Pre-set measurement time from 1 s to 24 hr in 1 s steps
Manual Controls: Reset, Start, Pause, Back-erase, Continue and Store
the measurement manually
BACK-ERASE
The last 5 s of data can be erased without resetting the measurement
PREFERENCES
Date, Time and Number formats can be specified
Measurement Status
ON SCREEN
Information such as overload and running/paused are displayed on
screen as icons
TRAFFIC LIGHTS
Red, yellow and green LEDs show measurement status and
instantaneous overload as follows:
• Yellow LED flash every 5 s = stopped, ready to measure
• Green LED flashing slowly = awaiting calibration signal
• Green LED on constantly = measuring
• Yellow LED flashing slowly = paused, measurement not stored
• Red LED flashing quickly = intermittent overload, calibration failed
NOTIFICATIONS
Sends an SMS or e-mail daily at a specified time or if an alarm condition
is fulfilled.
Alarm Conditions:
• Disk space below set value
• Internal battery enters set state
• Change in measurement state
• Reboot of instrument
Calibration
Initial calibration is stored for comparison with later calibrations
ACOUSTIC
Using Sound Calibrator Type 4231 or custom calibrator. The calibration
process automatically detects the calibration level when Sound Calibrator
Type 4231 is used
ELECTRICAL
Uses internally generated electrical signal combined with a typed-in value
of microphone sensitivity
CALIBRATION HISTORY
Up to 20 of the last calibrations made are listed and can be viewed on
the instrument
Signal Monitoring
The input signal can be monitored using an earphone/headphones
connected to the headphone socket
HEADPHONE SIGNAL
Input signal can be monitored using this socket with
headphones/earphones
Gain Adjustment: –60 dB to 60 dB
Start Time Stop Time Overload %
Elapsed Time L
Xeq
L
Yeq
L
XE
L
YE
L
Ceq
–L
Aeq
L
XSmax
L
XFmax
L
XImax
L
YSmax
L
YFmax
L
YImax
L
XSmin
L
XFmin
L
XImin
L
YSmin
L
YFmin
L
YImin
L
XIeq
L
YIeq
L
AIeq
–L
Aeq
L
AFTeq
L
AFTeq
–L
Aeq
Time Remaining
L
ep,d
L
ep,d,v
E
Dose% Proj. Dose% #VPeaks
(>NNNdB)
#VPeaks
(>137dB)
#VPeaks
(>135dB)
L
Vpeak
T
Vpeak
L
avUQ
TWA
TWA
v
DoseUQ% Proj. DoseUQ%
L
Aeq,T,mov,max
L
XS
L
XF
L
XI
L
YS
L
YF
L
YI
L
XS(SPL)
L
XF(SPL)
L
XI(SPL)
L
YS(SPL)
L
YF(SPL)
L
YI(SPL)
L
Vpeak,1s
L
AN1
or L
AUN1
L
AN2
or L
AUN2
L
AN3
or L
AUN3
L
AN4
or L
AUN4
L
AN5
or L
AUN5
L
AN6
or L
AUN6
L
AN7
or L
AUN7
Std.Dev.
L
Aeq,T,mov
Komentáře k této Příručce