Configure CV Parameters
There are several parameters that should be set up prior to operating in CV mode. Press
so that it lights up, then press and to access Setup for CV mode. The
setup menu will be shown:
The setup parameters are: Range, High (Current limit), and Low (Current limit). Use the
key to select each parameter, and use the numeric keypad to change the value. Press
to confirm the change.
Range
Use the numeric keypad to change the range. This value will also act as a limit to how much
voltage the load can be configured to. The adjustable ranges vary depending on the model.
See specifications for details.
High
This parameter refers to the current high limit for the automatic test mode. During automatic
test mode, the device under test (DUT) must be operating below the configured value for the
test to PASS upon completion. If the DUT operates above the configured value, the test will FAIL
upon completion.
Constant Voltage
Range=60.000V