7. Press and it will prompt to enter the slew rate of the step. If High-Rate was
selected earlier, units will be in A/us. Otherwise, it will be in A/ms.
8. Press and it will prompt to enter the step with. Valid range is 20 us 3600 s.
Enter a value and then press to continue.
9. The load will prompt to enter parameters for the next step. Follow steps 6-8 to enter
the values for the step’s parameters. Once all steps’ parameters are entered, the load
will prompt for the Repeat Count, which sets how many times to repeat the list
program. Valid range is 1 65535. Press to continue.
10. The load will prompt to select the memory location to store the list parameter
information. There are seven total save locations. Enter a value between 1 and 7, then
press to confirm. The display will return to the List menu.
Note: If List mode is enabled, the Edit menu will be locked and list parameters cannot be
changed until List mode is disabled.
EDIT LIST
Step 001 Width = 0.00002S
EDIT LIST
Step 001 Rate = 0.0001A/us
EDIT LIST
Step 001 Level = 0.0000A