B&K 2250 Light Uživatelský manuál Strana 131

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 212
  • Tabulka s obsahem
  • ŘEŠENÍ PROBLÉMŮ
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 130
CHAPTER 13
Tone Assessment Option BZ-7231 – 1/3-octave Method
123
Fig.13.6
The differences to the left
and right of the selected
frequency band are
shown in the tone
parameter panel – in this
example, there is a
11.4 dB difference to the
left and a 3.9 dB
difference to the right
The information shown in the value panel can be changed by tapping in the field. From the tone
parameters, the adjustment and the selected tone standard can be selected. See Fig.13.7.
Results from the tone assessment can also be found by tapping on the Broadband tab and
browsing the Broadband view in a similar way.
Fig.13.7
Changing the information
shown in the value panel
Zobrazit stránku 130
1 2 ... 126 127 128 129 130 131 132 133 134 135 136 ... 211 212

Komentáře k této Příručce

Žádné komentáře