APPENDIX A
Setup Parameters
143
A.5 Logged Broadband
T for LXeq,T,mov 00:01:00 to
01:00:00
Sets the averaging time T (in minutes) of the moving
L
Aeq
and moving L
Ceq
. When the averaging time T
has elapsed the L
Aeq,T,mov
and L
Ceq,T,mov
parameters will contain the L
Aeq
and L
Ceq
of the
latest T minutes – updated every second. In
addition, you also have
L
eq,T,mov
=L
Ceq,T,mov
–L
Aeq,T,mov
. The L
Xeq,T,mov
is
not stored with the total data, but can be logged
(BZ-7224 and BZ-7225). The L
Aeq,T,mov,max
,
L
Ceq,T,mov,max
and the difference between the C- and
A-weighted values for the complete measurement
period are stored.
Note: There are two sets of these parameters,
allowing two simultaneous moving averages: e.g.,
one with 15 min averaging time and one with 1 hour
averaging time.
Charge Injection Calibration
†
On
Off
Set to On for making a CIC at the start and end of
Logging. An Exclude marker is set on the profile and
Total parameters are not updated during the CIC
*. For Sound Level Meter (BZ-7130) and 1/1-octave and 1/3-octave Frequency Analyzer (BZ-7131 and BZ-7132) templates only
†. For Logging (BZ-7133) templates only
Table A.6 Logged broadband parameters
*
Parameter Values Comment
Full Statistics Yes
No
Determines whether the full broadband statistics
are logged or not
Broadband Parameters All
Selected
Determines whether all broadband parameters are
logged or a selected part is logged (up to 10
parameters)
Table A.5 (Cont.) Measurement control parameters