APPENDIX A
Setup Parameters
149
A.10 Occupational Health
Post-recording
Time
†
0 to 300 s Use this parameter to specify how much extra you want to be
recorded after the trigger conditions are no longer fulfilled
Duration Limit
†
On
Off
Use this parameter to enable the Minimum Duration and
Maximum Duration parameters for overruling the duration of
the signal recording determined by the trigger condition
parameters
Minimum Duration
†
00:00:00 to
01:00:00
When Duration Limit is On, then Minimum Duration will
determine the minimum recording time regardless of the
trigger conditions
The total record length will then (as a minimum) be the sum of
Minimum Duration, Pre-recording Time and Post-recording
Time
Maximum Duration
†
00:00:00 to
01:00:00
When Duration Limit is On, then Maximum Duration will
determine the maximum recording time regardless of the
trigger conditions.
The total record length will then (as a maximum) be the sum of
Maximum Duration, Pre-recording Time and Post-recording
Time.
Note: If Maximum Duration = 00:00:00, then it is disabled and
does not limit the duration
*. Requires license for Signal Recording Option BZ-7226.
†. For Logging BZ-7133 template only.
Table A.11 Occupational health parameters
Parameter Values Comment
Exposure Time 00:01:00 to
1.00:00:00
Set the Exposure Time to the actual time that
you are exposed to noise during a workday.
Used for calculation of L
ep,d
and L
ep,d,v
.
Note: Exposure Time can be changed after the
measurement has been done
Reference Time 00:01:00 to
5.00:00:00
Set the Reference Time as required for
calculation of Sound Exposure Level or Time
Weighed Average with a reference time other
than 8 hours.
Used for calculation of L
ep,d,v
and TWA
v
Table A.10 (Cont.) Signal Recording parameters
*