CHAPTER 12
Enhanced Logging (Optional Module)
133
Note: All other measured broadband and spectral parameters are logged automatically per
report period.
Then you can select:
Number of CICs
Desired time periods for calibration
6) Under the Logged Spectrum parameters, you can choose to log Full Spectral Statistics per
Logging Period or not.
Note: Logging the full spectral statistics at a short logging period (for example, every
second) takes up a lot of space on the memory card.
7) Under Triggers, you can specify a Signal Recording Timer for periodically starting a
signal recording (for example, five minutes per hour) and you can specify up to four
independent Level Triggers to be active at four different times during the day. Each level
trigger works as specified in section 11.1. In addition, you can specify a Hold Off period –
this means that during this period, new triggers are inhibited. You also have the ability to
trigger L
Xeq
(f1–f2).
Note: You can trigger on a specific frequency band by setting f1 = f2.
8) Under L
den
Periods parameters you can set up the day, evening and night periods and the
penalties in accordance with your local legislation.
To exit the screen, tap the icon.
12.2 Controlling the Measurement
The measurement is controlled in the same way you would control a normal Sound Level
Meter measurement, using Start/Pause , Continue , Back-erase and Reset
pushbuttons, see Chapter 3 for more details.
However, during the measurement, data are automatically saved in projects, one per day. The
projects are automatically saved at the time specified in the Save Project Data at parameter in
Setup > Measurement Control. The next project is (automatically) started immediately after the
first one, without any data loss.
Note: When Preset Logging Time has elapsed, or you pause the measurement, then the project
will not be saved automatically. This has to be done manually.
In the case of power failure, or other failures, the software is rebooted automatically. This
means that the measurement data collected up until the time of the reboot is saved in a project
and a new measurement is started. The new measurement is started in a new project.
The projects are named as specified in Preferences > Storage Settings.
Example: If you select Auto-naming of Projects, and start the measurement on the 13th of
November, then your projects for measuring 48 hours (assuming you have started in the middle
of the day, and have set Save Project Data at parameter to 00:00:00) will be:
051113 001 (containing data from the middle of the day November 13th until midnight)
Zobrazit stránku 144