B&K 2250 Light Uživatelský manuál Strana 146

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 332
  • Tabulka s obsahem
  • ŘEŠENÍ PROBLÉMŮ
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 145
134
Hand-held Analyzer Types 2250 and 2270 – User Manual
051114 001 (containing data for a whole day – November 14th)
051115 001 (containing data from midnight to the middle of the day on November 15th)
If you did not select Auto-naming of Projects then you would have the following projects
(assuming the Project Name Prefix parameter was set to Project and no projects had been saved
before in the directory):
Project 001
Project 002
Project 003
If you have set the Charge Injection Calibration parameter to On, then a CIC calibration is
performed at specified times of the day. An Exclude marker is set on the logging profile at the
point where the CIC is being performed. The update of the Total parameters and the Periodic
Reports is stopped while the CIC is being performed. You can see the results of the CIC
calibration under Total parameters, CIC Result 1 to 4.
The CIC result can be either ‘Passed’,Background noise too high’ or ‘CIC Ratio deviates
from Ref.’ (The ‘Ref.’ mentioned here is the reference created by the very first manual CIC, and
is the one used as the reference to all subsequent CIC ratio measurements, see section 5.10.)
Note: If a level trigger is active when the CIC is due to start, then the CIC will not be done.
12.2.1 Annotating a Project
Measurements can be annotated in the same way as you annotate Logging measurements. See
section 11.2.1.
12.2.2 Recording the Signal
Signal recordings can be done in the same way as described under Logging measurements. See
section 11.2.2.
12.3 Displaying the Results
The Enhanced Logging Measurement screen includes three tabs at the bottom: Profile,
Spectrum and Broadband view – as the Logging Measurement screen does. You view data in
the same way as in the Logging module. You mark and annotate sound categories and edit
markers and annotations in the same way as in the Logging module – please see Chapter 11.
The Result Selector (first item in the third line of the Status Field) determines whether results
from the Total, Logged, Logged (100 ms) or Report measurement are displayed.
If you need to display data from another day of the measurement, then use Explorer to select
and open the project from that day.
Zobrazit stránku 145
1 2 ... 141 142 143 144 145 146 147 148 149 150 151 ... 331 332

Komentáře k této Příručce

Žádné komentáře

Clickfree Traveler podręczniki

Instrukcje obsługi i podręczniki użytkownika dla Sprzęt komputerowy Clickfree Traveler.
Dostarczamy 1 pdf podręczników Clickfree Traveler do pobrania za darmo według typów dokumentów: Instrukcja Użytkownika


Clickfree Traveler Instrukcja Użytkownika (2 strony)


Marka: Clickfree | Kategoria: Hardware | Rozmiar: 0.09 MB |

 

Spis treści





Więcej produktów i instrukcji dla Sprzęt komputerowy Clickfree

Modele Rodzaj dokumentu
DVD Transformer Instrukcja Użytkownika   Clickfree DVD Transformer User Manual, 37 strony
Transformer SE Instrukcja Użytkownika   Clickfree Transformer SE User Manual, 132 strony
5QSG Instrukcja Użytkownika   Clickfree 5QSG User Manual, 2 strony
C6 Instrukcja Użytkownika   Clickfree C6 User Manual, 24 strony
C2 Instrukcja Użytkownika   Clickfree C2 User Manual, 76 strony
Backup Drive Instrukcja Użytkownika   Clickfree Backup Drive User Manual, 33 strony