CHAPTER 17
Low Frequency Option BZ-7234
195
10) Select Units > Engineering Unit and select No (in dB) or Yes (in m/s
2
).
Note: If you want readouts in g instead of m/s
2
, set the Vibration Unit to US/UK in
Preferences > Regional Settings.
11) Tap to exit Setup.
17.4.1 Controlling the Measurement
The measurement is controlled in the same way you would control a normal Sound Level
Meter measurement, using Start/Pause , Continue , Back-erase and Reset
pushbuttons, see Chapter 3 for more details.
17.4.2 Displaying the Results
Tap on the broadband result readouts to select parameters from the Low Frequency group to
select the building vibration results (Fig.17.6).
Fig.17.6
Selection of Low
Frequency building
vibration results
17.5 Setting Up the Frequency Analyzer Module for Very
Low Frequencies
1) For:
– Sound measurements: Mount your low-frequency microphone and low-frequency
adaptor on the preamplifier and connect it to the top socket (see section 2.5.2). If it is
your first time to use the microphone, it needs to be created in the transducer database
(see section 5.8) and calibrated (see section 5.3)
– Vibration measurements: Connect the accelerometer to the rear socket. Brüel & Kjær
recommends Accelerometer Type 8344 for low-level building vibration
measurements. If it is your first time to use the microphone, it needs to be created in
the transducer database (see section 5.8) and calibrated (see section 5.3)