CHAPTER 16
2-channel Option BZ-7229
185
When the same type of transducers are used by both channels (e.g., two microphones or two
accelerometers) or if you only display data from one channel, then the spectrum display is the
standard spectrum display (Fig.16.7).
Fig.16.7
Single-channel spectrum
When the transducers at the two channels are of different types (e.g., a microphone and an
accelerometer), the scaling of the two spectra can be set individually for easy comparison. Ch.1
will use the standard Y-axis at the left of the spectrum and Ch.2 will have its own Y-axis to the
right of the spectrum (instead of the broadband values). Tapping on the Ch.2 Y-axis reveals the
same possibilities for zooming and scaling the spectrum as for Ch.1. See examples in Fig.16.8.
Fig.16.8
Displaying the results with different scaling for each channel
16.5.3 Saving Results
Measurements are saved and can be opened later in the same way as described for the Sound
Level Meter project in Chapter 3.