APPENDIX A
Setup Parameters
249
A.18 Post-processing
Build-up Time 1 to 20 s Set the Build-up Time to allow the sound
pressure in the room to settle after the sound
source is switched on
Bottom Frequency
50 Hz to Top Frequency 1/1-octave: 63 Hz – 8 kHz
1/3-octave: 50 Hz – 10 kHz
Note: The settings of Bottom and Top
Frequency control the bandwidth of the noise
from the internal noise generator
Top Frequency
Bottom Frequency to
10 kHz
1/1-octave: 63 Hz – 8 kHz
1/3-octave: 50 Hz – 10 kHz
*. Available for Frequency Analysis Software BZ-7223, single channel, and Reverberation Time Software BZ-7227 and
microphone input only.
†. Available for Frequency Analysis Software BZ-7223 only.
Table A.18 Post-processing Parameters
*
Parameter Values Comment
Ensemble Averaging No
Yes
Set to Yes to average each of the position
decays into an average decay (called the
ensemble average or room average). The
averaged decays of the room can then be
displayed in the Decay View. T30 Room, T20
Room and EDT Room will be calculated from
the room average.
Set to No for no ensemble averaging. There
will be no decays available for the room. T30
Room, T20 Room and EDT Room will be
calculated as averages of the T30, T20 and
EDT, respectively, for all the positions
b
Table A.17 (Cont.) Generator Parameters
*
Parameter Values Comment